|
It was reported that on (b)(6) 2026, according to the standard surgical procedure, the patient was scheduled to undergo hysteroscopic endometrial lesion resection + hysteroscopic cervical lesion resection + fractional diagnostic curettage + cervical biopsy.During the operation, while using the high-frequency electrosurgical device, the working handpiece suddenly burned through the bipolar electrode cable, causing the patient to experience a brief electric shock.The use of the device was immediately discontinued, and a backup working handpiece and electrode cable were replaced.After confirming no issues, the surgery continued.New additional information was received on july 24, 2026.Based on the new information there are 3 additional potentially causal karl storz products involved in the event and additional reported error description: "the sealing ring leaks, and water flows through the gap to the electrode cable during the procedure." cross reference mdr: 9610617-2026-01599, 9610617-2026-01797.9610617-2026-01798.
|