• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: KARL STORZ SE & Co. KG Working Element, bipolar

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

KARL STORZ SE & Co. KG Working Element, bipolar Back to Search Results
Model Number 26040EB
Medical Device Problem Code Insufficient Device Problem Information (3190)
Health Effect - Clinical Code Electric Shock (2554)
Date of Event 06/18/2026
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that on (b)(6) 2026, according to the standard surgical procedure, the patient was scheduled to undergo hysteroscopic endometrial lesion resection + hysteroscopic cervical lesion resection + fractional diagnostic curettage + cervical biopsy.During the operation, while using the high-frequency electrosurgical device, the working handpiece suddenly burned through the bipolar electrode cable, causing the patient to experience a brief electric shock.The use of the device was immediately discontinued, and a backup working handpiece and electrode cable were replaced.After confirming no issues, the surgery continued.New additional information was received on july 24, 2026.Based on the new information there are 3 additional potentially causal karl storz products involved in the event and additional reported error description: "the sealing ring leaks, and water flows through the gap to the electrode cable during the procedure." cross reference mdr: 9610617-2026-01599, 9610617-2026-01797.9610617-2026-01798.
 
Additional Manufacturer Narrative
Under user's discretion the device in question will not be returned to the manufacturer for evaluation.The investigation will be performed by a designated karl storz employee.The event is filed under internal karl storz complaint id: (b)(4).Cross reference internal complaint id: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
Working Element, bipolar
Common Device Name
Working Element, bipolar
Manufacturer (Section D)
KARL STORZ SE & Co. KG
dr.-karl-storz-strasse 34
tuttlingen 78532
GM  78532
MDR Report Key50122169
Report Number50122169
Device Sequence Number10560385
Product Code HIH
UDI-Device Identifier04048551091784
UDI-Public04048551091784
Combination Product (Y/N)N
Initial Reporter CountryCH
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 08/03/2026
Report Date (Section F) 08/03/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number26040EB
Device Catalogue Number26040EB
Device Lot NumberUnknown
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Distributor Facility Aware Date07/24/2026
Event Location Hospital
Date Report to Manufacturer08/03/2026
Initial Date Received by Manufacturer Not provided
Initial Report FDA Received Date08/03/2026
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-