• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: Cardiovascular Systems Incorporated Diamondback Coronary; Catheter, Coronary, Atherectomy

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Cardiovascular Systems Incorporated Diamondback Coronary; Catheter, Coronary, Atherectomy Back to Search Results
Model Number DBEC-125
Medical Device Problem Code Entrapment of Device (1212)
Health Effect - Clinical Code Device Embedded In Tissue or Plaque (3165)
Date of Event 09/07/2025
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that a diamondback 360 coronary orbital atherectomy device (oad) was used to treat the left main and circumflex arteries.Upon removal, the crown became stuck in the vessel.A secondary wire and balloon were used to free the crown from the vessel.There was no clinically significant delay to the procedure and no patient adverse effects.There is no additional information.
 
Event or Problem Description
It was reported that a diamondback 360 coronary orbital atherectomy device (oad) was used to treat the left main and circumflex arteries.Upon removal, the crown became stuck in the vessel.A secondary wire and balloon were used to free the crown from the vessel.There was no clinically significant delay to the procedure and no patient adverse effects.There is no additional information.
 
Additional Manufacturer Narrative
Additional information: d9, h3, h6 (component code, type of investigation, investigation findings, investigation conclusions) , h11.A visual inspection, functional testing, and detailed analysis were performed on the returned device.The reported driveshaft fracture was confirmed.Production record and corrective and preventative actions (capa) reviews were performed and revealed no indication of a product quality issue.Additionally, a query of the complaint handling database for the reported lot revealed there is no indication of a lot specific issue.Based on the reported information and the observations from the returned analysis, the investigation determined that the driveshaft fracture appears to be related to circumstances of the procedure.Detailed analysis of the driveshaft fracture revealed evidence indicating that the fracture occurred while spinning in an elevated stress environment.The cause of the stress condition that led to the fracture is unknown.Based on the results of the complaint investigation there is no indication of a product quality issue with respect to the design, manufacture, or labeling of the device.
 
Additional Manufacturer Narrative
Manufacturer's investigation is still pending at this time.Results and conclusions will be provided in the final report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
Diamondback Coronary
Common Device Name
Catheter, Coronary, Atherectomy
Manufacturer (Section D)
Cardiovascular Systems Incorporated
1225 old highway 8 nw
st. paul MN 55112
Manufacturer (Section G)
Cardiovascular Systems Incorporated, #3004742232
1225 old highway 8 nw
saint paul MN 55112
Manufacturer Contact
alyssa mickelson
1225 old highway 8 nw
st. paul, MN 55112
+1(651)756-2718
MDR Report Key50122270
Report Number3004742232-2026-00193
Device Sequence Number5584674
Product Code MCX
UDI-Device Identifier10850026568698
UDI-Public(01)10850026568698(17)271231(10)594732-1
Combination Product (Y/N)N
Initial Reporter StatePR
Initial Reporter CountryUS
PMA/510(K) Number
P130005
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 10/01/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberDBEC-125
Device Catalogue Number7-10098-01
Device Lot Number594732-1
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer08/05/2026
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/09/2026
Supplement Date Received by Manufacturer09/08/2026
Initial Report FDA Received Date08/03/2026
Supplement Report FDA Received Date10/01/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured12/05/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-