• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZOLL MEDICAL CORPORATION X SERIES; AUTOMATED EXTERNAL DEFIBRILLATORS (NON-WEARABLE)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZOLL MEDICAL CORPORATION X SERIES; AUTOMATED EXTERNAL DEFIBRILLATORS (NON-WEARABLE) Back to Search Results
Model Number 620-2231411-01
Medical Device Problem Codes Failure to Charge (1085); Failure to Deliver Shock/Stimulation (1133)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 07/09/2026
Type of Reportable Event Death
Event or Problem Description
Complainant alleged while attempting to treat a patient (age & gender unknown), the device failed to discharge.Complainant indicated that the clinician obtained another device to continue treating the patient.Complainant indicated that the patient subsequently expired.
 
Event or Problem Description
Complainant alleged while attempting to treat a patient (age & gender unknown), the device failed to charge.
 
Additional Manufacturer Narrative
Zoll medical corporation has not received the device for evaluation and this complaint is still under investigation.
 
Additional Manufacturer Narrative
This supplemental report is reporting the evaluation of the device and correcting information submitted on the initial report.Please reference b5 and h6: incident description, medical device problem code the device was returned to zoll medical corporation for evaluation.The customer's report was observed during review of the device data logs.However, the device was put through extensive testing including shock testing and all functional testing without duplicating the report.The customer did not return any accessories, including the battery, which is suspected to have been low of charge during the event.The x series operator's guide speaks to proper battery management, instructing the user to, "always carry at least one fully charged spare battery.If no other source of back-up power is available, two spare batteries are advisable." the device was recertified and returned to the customer.Analysis of reports of this type has not identified an increase in trend.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
X SERIES
Common Device Name
AUTOMATED EXTERNAL DEFIBRILLATORS (NON-WEARABLE)
Manufacturer (Section D)
ZOLL MEDICAL CORPORATION
269 mill road
chelmsford MA 01824
Manufacturer Contact
scott august
269 mill road
chelmsford, MA 01824
+1(978)421-9552
MDR Report Key50122380
Report Number1220908-2026-02895
Device Sequence Number10527500
Product Code MKJ
UDI-Device Identifier00847946019273
UDI-Public(01)00847946019273
Combination Product (Y/N)N
Initial Reporter StateMO
Initial Reporter CountryUS
PMA/510(K) Number
P160022
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Type of Report Initial,Followup
Report Date (Section B) 07/13/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Other
Device Model Number620-2231411-01
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer08/10/2026
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/13/2026
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date08/03/2026
Supplement Report FDA Received Date09/14/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Outcome Attributed to Adverse Event Death;
Patient Age71 YR
Patient SexMale
-
-