It was reported that the patient with this inflatable penile prosthesis (ipp) was unable to inflate the device.During revision surgery, no fluid was observed moving into the cylinders.The device was disconnected from the reservoir, and surrogate testing was performed, after which the cylinders inflated.An attempt was made to aspirate fluid from the reservoir; however, no fluid could be retrieved.As a result, the reservoir was capped and retained in situ, and a new reservoir was implanted on the left side.Following implantation of the new reservoir, the device inflated properly.No patient complications were reported.
|
|
The device is not available for analysis; therefore, no physical analysis of the product could be performed.The reported device performance allegation cannot be confirmed.Based on the information available the cause that contributed to the reported event cannot be established.Therefore, a conclusion code of cause not established was assigned to this investigation.Model number/catalog number: 72404433.Serial number: n/a.Batch/lot number: 1100767343.Model/catalog description: cx preconnect tenacio 21cm ps, iz.Unique identifier (udi) #: (b)(4).
|