Supplemental report 01.Unomedical hereby submits this supplemental report as part of its complaint remediation activities conducted under a capa 2356421 and capa 2566479 in accordance with the fda action plan, which incorporates qp-ic-2026-0020 (ic retrospective complaint review) and qp-ic-2026-0024 (ic retrospective mdr review).This submission includes a retrospective reassessment of previously evaluated complaints.Unomedical is providing this supplemental information in accordance with the reporting requirements set forth in 21 cfr part 803.Any fields left blank indicate that the information is unknown, unavailable, or remains unchanged.Correction: this mdr is being submitted to correct the imdrf components code under h6
additional information -this electronic medical device report (emdr) is being submitted to include the below:
h6: investigation results under type of investigation, investigation findings, investigation conclusions.H11: investigation summary.Complaint investigation results:
review of complaint record event description and all available information was completed and no lot number specific information identified.Requests for lot specific information were sent to the customer but were unable to be provided.As a result, an investigation was completed at the product family level inset ii and malfunction type leakage.Corrective and preventive action (capa) determination:
based on the available information, no further investigation is required, nor capa plan is needed.The record will be closed and monitored through tracking and trending (monthly trips and alerts).Summary conclusion:
based on the available information and the investigation performed, the complaint will be closed as complaint unconfirmed due to lack of information.Based on the available information, this event is deemed to be a serious injury.Request was performed for additional information including lot number; however, lot number was not provided.A complaint investigation was initiated under complaint investigation.Unfortunately, no specific lot number was identified, which limits the ability to trace or analyze a particular item.To date no additional information has been received.Should additional information become available, a follow-up report will be submitted.Fda registration number
reporting site: 8021545.Manufacturing site: 3003442380.
|
|
Mdr 3003442380-2026-64340 / initial 5 of 6.Unomedical hereby submits this initial report as part of its complaint remediation activities conducted under a capa 2356421 and capa 2566479 in accordance with the fda action plan, which incorporates qp-ic-2026-0020 (ic retrospective complaint review) and qp-ic-2026-0024 (ic retrospective mdr review).This submission includes a retrospective reassessment of previously evaluated complaints.Unomedical is providing this initial information in accordance with the reporting requirements set forth in 21 cfr part 803.Any fields left blank indicate that the information is unknown, unavailable, or remains unchanged.E1: name: (b)(4).Based on the available information, this event is deemed to be serious injury.To date no additional information has been received.Should additional information become available, a follow-up report will be submitted.Fda registration number.Reporting site: 8021545.Manufacturing site: 3003442380.
|