Block h6:
imdrf device code a0401 captures the reportable event of stent shaft break.Block h11:
the polaris ultra was returned for analysis.During the inspection no damage was found in the returned device not being able to confirm the reported issue.Based on the most relevant information of the complaint event description, device analysis and including the product record review results, the device meets all manufacturing specifications required and passed all the controls and inspections, no abnormalities were reported during the assembly process.This complaint is assigned an investigation conclusion code of no problem detected.A review of the device history record (dhr) was confirmed this device met specification prior to distribution and there were no manufacturing deviations which could have contributed to the reported event.A risk review was completed and confirmed that the event of stent shaft break was defined in the risk documentation and is documented accordingly.This event type has been accounted during product risk analysis to support acceptable risk benefits for the product.A similar complaint review was completed using global similar complaint review tool, which reviewed ar code stent shaft break and did not identify any emerging trends or similar complaints that require escalation.Therefore, this conclusion was selected because it was reported stent shaft break, however, after visual inspections in the complaint device, it was not possible to identify any evidence of either the alleged issue or any defect on the device.
|