a getinge field service engineer (fse) inspected the unit and determined that the failure was associated with the drive assembly.The equipment had exceeded the recommended service interval; therefore, replacement of the 5,000 hour preventive maintenance (pm) kit was also recommended.The device was removed from service.During the repair, the fse replaced the drive assembly and the 5,000 hour pm kit.Following the repair, all functional tests were performed and found satisfactory.The unit was monitored for approximately two hours, during which no recurrence of the low vacuum error was observed.The unit was subsequently returned to the customer in good working condition.
|