Zimmer Biomet, Inc. G7 DUAL MOBILITY LINER 38MM C; Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Porous Uncemented
|
Back to Search Results |
|
| Catalog Number |
110024461 |
| Medical Device Problem Codes |
Detachment of Device or Device Component (2907); Noise, Audible (3273)
|
| Health Effect - Clinical Codes |
Failure of Implant (1924); Pain (1994)
|
| Date of Event |
07/17/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that the patient required an acetabular revision approximately two months post-implantation due to dissociation of the dual-mobility components, specifically the ceramic head and polyethylene liner.The patient reported an internal rotation movement of the limb while rising from an armchair.This resulted in a dislocation of the prosthesis, which was reduced in the emergency department.In reality, dissociation of the dual-mobility (dm) component was already present but went unnoticed in the ed; consequently, the patient consulted a doctor due to pain and audible noise from the prosthesis.Follow-up x-rays and ct scans revealed that the dm cup was displaced
this necessitated a revision procedure for the component.Attempts have been made and no further information has been provided.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).D10: zimmer biomet part number 00877502802 bioloxa delta, ceramic femoral head, m, a¸ 28/0, taper 12/14 lot number 3252839.G2: foreign - event occurred in italy.The customer has indicated that the product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.
|
| |
|