| Model Number |
LAS0036 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/27/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
The user facility reported that during a patient procedure the light handle cover to their harmonyair a-series surgical lighting system detached from the light handle; the cover did not enter the sterile field.The procedure was completed successfully.No report of injury.
|
| |
|
Additional Manufacturer Narrative
|
|
The device subject of the reported event was not returned for evaluation.Without the return of the device, a cause of the reported event could not be determined.Based on the description of the reported event, the likely cause of the reported detachment would be attributed to incorrect or incomplete installation of the cover by user facility personnel.The recommended instructions for installation of the las0036 light handle cover product is documented within the applicable steris surgical lighting system's operator manual.The relevant operator manual is provided with every steris surgical lighting system sold.A steris account manager provided in-service training on the proper installation and handling of the light handle cover.No additional issues have been reported.
|
| |
|
Search Alerts/Recalls
|