| Model Number |
M00553640 |
| Medical Device Problem Codes |
Use of Device Problem (1670); Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Ulcer (2274)
|
| Date of Event |
07/13/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported to boston scientific corporation that an axios stent and electrocautery enhanced delivery system was intended to be implanted in the duodenal bulb to treat acute cholecystitis during an endoscopic ultrasound (eus) procedure performed on (b)(6) 2026.It was reported that the patient had an axios stent implanted and a cook 7 fr x 4cm double pigtail through stent went through the axios stent.The patient came back for a follow up and axios removed with a rat tooth forceps when the physician noticed that there was severe ulceration around the contralateral wall of duodenum.The patient was treated with high dose of ppi (proton pump inhibitor).It was also reported that the patient was not suitable for laparoscopic cholecystectomy due to many underlying conditions, so the gi was consulted for axios placement.There were no patient complications reported as a result of this event at this time.Note: it was reported that the axios stent was intended to be implanted during a pancreatic cystic gastrectomy with gallbladder drainage - acute cholecystitis procedure.However, according to the axios stent and electrocautery-enhanced delivery system instructions for use, the stent is indicated for use to facilitate transgastric or transduodenal endoscopic drainage of symptomatic pancreatic pseudocysts 6 cm or larger in size and walled-off necrosis 6 cm or larger in size with at least 70% fluid content, provided they are adherent to the gastric or bowel wall.The device is not indicated for a gallbladder drainage - acute cholecystitis.
|
| |
|
Additional Manufacturer Narrative
|
|
Block d4, h4: the complainant was unable to provide the complaint device lot number.We are unable to provide the device manufacture date.If additional details become available, a supplemental report will be submitted.Block h6: imdrf impact code f2301 captures the reportable event of additional device required.Imdrf impact code f2303 captures the reportable event of medication required.Imdrf patient code f2303 captures the reportable event of ulcer.
|
| |
|
Search Alerts/Recalls
|