• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: Physio-Control, Inc. - 3015876 LIFEPAK(R) 1000 DEFIBRILLATOR; Automated External Defibrillators (Non-Wearable)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Physio-Control, Inc. - 3015876 LIFEPAK(R) 1000 DEFIBRILLATOR; Automated External Defibrillators (Non-Wearable) Back to Search Results
Model Number 1000
Medical Device Problem Code Delayed Charge Time (2586)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/15/2026
Type of Reportable Event Malfunction
Event or Problem Description
During preventive maintenance, stryker observed that the device will not pass the shock time limit test and that defibrillation charge is delayed.In this state the device may not be able to deliver defibrillation therapy, if needed.There was no patient use associated with the reported event.
 
Additional Manufacturer Narrative
The customer's device was not returned for further evaluation.A cause of the reported issue could not be determined.The customer received a replacement device.A quality engineer confirmed with the customer that the device had been scrapped.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIFEPAK(R) 1000 DEFIBRILLATOR
Common Device Name
Automated External Defibrillators (Non-Wearable)
Manufacturer (Section D)
Physio-Control, Inc. - 3015876
11811 willows road ne
redmond WA 98052
Manufacturer (Section G)
Physio-Control, Inc. - 3015876
11811 willows road ne
redmond WA 98052
Manufacturer Contact
juan salazar
11811 willows road ne
redmond, WA 98052
+1(425)867-4000
MDR Report Key50129281
Report Number0003015876-2026-01569
Device Sequence Number10527710
Product Code MKJ
UDI-Device Identifier00883873815458
UDI-Public00883873815458
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
K122600
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source User Facility
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 09/23/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model Number1000
Device Catalogue Number99425-000023
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer07/15/2026
Is the Reporter a Health Professional? No
Initial Date Received by Manufacturer 07/15/2026
Supplement Date Received by Manufacturer09/21/2026
Initial Report FDA Received Date08/04/2026
Supplement Report FDA Received Date09/23/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/01/2011
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-