| Model Number |
1201 |
| Medical Device Problem Codes |
High impedance (1291); Communication or Transmission Problem (2896)
|
| Health Effect - Clinical Codes |
Micturition Urgency (1871); Urinary Frequency / Polyuria (2275); Urinary Incontinence (4572); No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/01/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that the patient with an snm tined lead had symptoms previously well-controlled that gradually got worse over 2-3 weeks.On checking their remote, the patient noticed a red error light, which could not be cleared over the phone.In clinic, the clinician programmer displayed 3 electrode open impedances and programming around the errors temporarily was performed, but patient symptoms did not improve.X-rays were taken with no issues found.The patient reported having no recent falls or trauma only bumped the back of their head lightly with no suspected damage.Patient underwent a successful full revision.They reported feeling well with bowel and bladder symptoms improved and no error codes.No other patient complications were reported as a result of this event.
|
| |
|
Additional Manufacturer Narrative
|
|
Block d2b: product code - additional product code qon.Block g4: premarket/510(k) # - additional premarket/510(k) p190006.
|
| |
|