|
Brazil.It was alleged that approximately three months after surgery, the patient developed a deformity of the left breast in the upper outer quadrant.It was further alleged that asymmetry of the inframammary fold was observed, prompting a bilateral breast mri.Left breast implant was found to be ruptured.The investigation is ongoing.
|
|
As stated in the literature, ¿a number of risk factors for rupture have been identified; the most common cause is surgical instrument damage.¿ (handel et al., 2013).Additionally, rupture is a common and known reason for complaints, and it is clearly characterized in the product dfu included with the implant, as stated above.Additionally, no cases of rupture because of product failure have been reported to establishment labs ever.Per our directions for use, each patient must receive the establishment labs s.A.¿motiva implants®: information for the patient¿ during her surgical consultation, it is the surgeons´ responsibility to ensure that the patient completely understands the information regarding the risks, benefits, and recommendations associated with silicone gel-filled breast implant surgery, as well as the complications typical of any type of surgery.This document is available in motiva® website: ifu.Motiva.Health
the instructions for use in the directions for use were reviewed, to determinate if there are indications that ensure the prevention and good handling of the product.For this event, it is considered that the information is suitable for the section of surgical precautions as follows:
¿breast implants can potentially remain intact for decades in the body, but all such devices will fail at some point.Breast implants rupture when the shell develops a tear or hole.Rupture can occur at any time after implantation, but it is most likely to occur; the longer the implant is in place.The following may cause implants to rupture: damage by surgical instruments, implant stress, and weakening during implantation, age and design of the implant, submuscular rather than sub-glandular location, the occurrence of post-operatory hematomas or seromas, folding or wrinkling of the implant shell, excessive force to the chest, trauma, compression during mammographic imaging, and severe capsular contracture¿.Establishment labs requests the unit to confirm the event and to perform the corresponding testing (including but not limited to: revision and characterization under the microscope of the rupture section, mechanical testing according to approved standards, etc.).For this specific case, no response was received regarding device return.Breast implants are not considered lifetime devices, and implant rupture is a risk associated with breast surgery, which can occur at any time.
|