| Model Number |
D433 |
| Medical Device Problem Codes |
Pacing Problem (1439); Interrogation Problem (4017)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/27/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that the physician was unable to interrogate this implantable cardioverter defibrillator (icd) device post an ablation procedure.The physician reported attempting to interrogate with multiple programmers and troubleshooting with magnets but was unable to interrogate the device.The device was programmed to ventricular pacing with no evidence of pacing but patient has intrinsic rhythm.Technical services (ts) was consulted for troubleshooting assistance.Additional troubleshooting attempts were unsuccessful.Subsequently, a device replacement procedure was performed, this icd was explanted and replaced with a new device successfully.No additional adverse patient effects were reported.This icd was returned to facility and is awaiting analysis.
|
| |
|
Additional Manufacturer Narrative
|
|
This product is being evaluated in our post market quality assurance laboratory.This report will be updated when evaluation is complete.
|
| |
|