| Model Number |
DDMB1D4 |
| Medical Device Problem Code |
Failure to Deliver Shock/Stimulation (1133)
|
| Health Effect - Clinical Code |
Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
|
| Date of Event |
07/15/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
It was reported by a nurse that the implantable cardioverter defibrillator (icd) failed to treat an episode of ventricular tachycard ia (vt).It was noted the patient was unresponsive during the episode and required external defibrillation.Upon technical review, it was noted that the ventricular rate during the episode fell into the monitor only zone, so it did not meeting rate zone criteria for therapy.The icd remains in use.No further patient complications have been reported as a result of this event.
|
| |
|