| Model Number |
MAJ-210 |
| Medical Device Problem Codes |
Break (1069); Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/06/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
No additional information received from customer.
|
| |
|
Event or Problem Description
|
|
It was reported during a diagnostic endobronchial ultrasound with guide sheath (ebus-gs) procedure there was a black foreign material found inside the patient body.It was through to be a damaged part of the single use biopsy valve.It was suctioned and retrieved.The procedure was prolonged less than thirty minutes.There were no reports of patient harm.
|
| |
|
Additional Manufacturer Narrative
|
|
This supplemental report is being submitted to provide the results of the final investigation.Updated fields: d9, h2, h3, h6, h7, h9, h11.The device was returned to olympus for inspection, and the reported failure was confirmed.The device evaluation found that part of the biopsy valve was broken and a broken piece of the biopsy valve fell off.It was confirmed that the broken piece had been retrieved as the shape of the detached rubber fragment matches that of the damaged portion of the biopsy valve, leading to the conclusion that the detached rubber fragment is part of the biopsy valve.A root cause could not be identified.Based on the results of the investigation, it is likely the following led to the malfunction: it is possible that the biopsy valve broke due to stress caused by inserting or removing the syringe at an angle.Should additional relevant information become available, a supplemental report will be submitted.Olympus will continue to monitor field performance for this device.
|
| |
|
Additional Manufacturer Narrative
|
|
E1/initial reporter establishment name: (b)(6)
the investigation is ongoing.A supplemental report will be submitted when the investigation is completed or if additional information becomes available.
|
| |
|