| Model Number |
PT-000664 |
| Medical Device Problem Code |
Use of Incorrect Control/Treatment Settings (1126)
|
| Health Effect - Clinical Codes |
Hypoglycemia (1912); Malaise (2359); Syncope/Fainting (4411)
|
| Date of Event |
07/04/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that the patient visited the emergency room with hypoglycemia on (b)(6) 2026.The patient's blood glucose levels declined to 50 mg/dl while wearing the pod(leg).The patient stated that they had transferred omnipod 5 therapy settings to a new phone application and may have entered settings incorrectly during the phone change, which resulted in receiving more insulin than expected.Symptoms reported include malaise and fainting.The patient was treated with intravenous (iv) fluids and was released the same day.The pod was discarded.
|
| |
|
Additional Manufacturer Narrative
|
|
According to the complainant the device will not be available for investigation because it was discarded.We are unable to determine if any product condition could have contributed to the reported event.Lot release records were reviewed, and the product lot met all acceptance criteria.Locked down smartphone: phone_control_ios.Omnipod software app version: 3.1.6.Operating system: n5004l-am-q-mv01602-06-01.06.Hardware: n5004l.Cgm sensor type: g7.Please note, the device identifiers are captured as reported by the complainant and may not align with the device configuration reported in this section as this data is pulled from our cloud based on the reported date of event.
|
| |
|