It was reported that the patient with this implantable neurostimulator (ins) experienced an increase in fecal incontinence.When attempting to increase stimulation, a red indicator light was observed on the remote.The ins was charged; however, the red indicator light persisted.Troubleshooting was performed over the phone, including switching between programs, without resolution, as a continuous red indicator light remained present on both programs.No recent falls were reported.An appointment was scheduled for further evaluation and in-person troubleshooting with the clinical specialist.Additional information received indicated that evaluation identified a fractured lead.The patient subsequently underwent explant and reimplant of the system.No further patient complications were reported.
|