| Model Number |
71020 |
| Medical Device Problem Code |
Intermittent Communication Failure (4038)
|
| Health Effect - Clinical Code |
Inadequate Pain Relief (2388)
|
| Date of Event |
07/20/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
The patient was implanted with a nalu peripheral nerve stimulator system on (b)(6) 2026 targeting the cluneal nerve.After using the system for some time the patient then reported challenges connecting the implantable pulse generator (ipg) to the external therapy discs.Troubleshooting in the field confirmed the external devices to be functioning as expected and found no impedance errors when the system was tested while implanted.Fluoroscopic imaging was completed to rule out ipg migration as a cause of the connectivity issues.Cause of the connectivity issues was unable to be identified so the patient was scheduled for revision of the ipg.On (b)(6) 2026, a surgical procedure was performed to reposition the ipg, however after handling the device it returned impedance errors and required replacement during the same procedure.The implanted leads remained in place and were successfully connected to the new ipg.
|
| |
|
Additional Manufacturer Narrative
|
|
The nalu representative present at the revision confirms there are no reports of any events leading up to the change in connectivity and imaging did not identify any implanted device migration.Impedance testing returned all good readings before revision and then impedance errors after removing the ipg from the pocket.There is insufficient information or testing to verify if the impedance errors were due to device failure or related to handling damage from the revision procedure.The explanted ipg was not retained for return to the firm and thus no further testing was able to be performed to draw any conclusions as to the cause of the connectivity challenges.
|
| |
|
Search Alerts/Recalls
|