| Model Number |
Mi1250 SYNCHRONY 2 |
| Medical Device Problem Code |
Migration (4003)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/31/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
The hearing performance decreased and imaging confirmed an electrode extrusion.Re-implantation will be done on (b)(6)2026.
|
| |
|
Event or Problem Description
|
|
The hearing performance decreased and imaging confirmed an electrode extrusion.The user was reimplanted.
|
| |
|
Additional Manufacturer Narrative
|
|
The device has not been explanted.If it should be explanted, it should be returned to the manufacturer for evaluation.When available, a device failure analysis will be submitted as a follow up report.
|
| |
|
Additional Manufacturer Narrative
|
|
Conclusion: device investigations did not reveal any device defect or problem which is expected to have been present whilst implanted.Mechanical damages found during investigation are attributable to the removal surgery.The device was not providing benefit to the patient due to an active electrode migration out of the cochlea.This is a final report.
|
| |
|