(b)(4).Updated sections: b4, g3, g6, h2, h3, h6, h11.The device was returned to the factory for evaluation on 07/20/2026.An investigation was conducted on 8/18/2026.A visual inspection was conducted.Signs of clinical use and no evidence of blood was observed.The delivery device and the intact seal and tension spring was returned outside the loading device with the white plunger not depressed and the blue safety lock on, which prevents the white plunger from being depressed.There were no visual defects observed on the intact seal or tension spring assembly.Measurements of the delivery device were taken; the inner diameter was measured at 0.195 inches; the outer diameter was measured at 0.219 inches (rm2036883).The length of the delivery tube was measured at 2.49 inches (mcv00004217).The measurement values recorded for the delivery tube were within the tolerance specifications.Based on the returned condition of the device as well as the evaluation results, the reported failures "fitting problem" and "detachment of device or device component" were not confirmed.The lot #3000500748 history record review was completed.There was an ncmr, rework, or deviation documented for the reported lot number.Based on the dhr/lhr review results, it was determined that there is a relation between the batch manufacturing process and the reported failures.
|