Maquet Cardiovascular LLC HST III System (3.8MM); Clamp, Vascular Back to Search Results
Model Number HSK-3038
Medical Device Problem Codes Fitting Problem (2183); Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/13/2026
Type of Reportable Event Malfunction
Event or Problem Description
N/a.
 
Additional Manufacturer Narrative
(b)(4).Updated sections: b4, g3, g6, h2, h3, h6, h11.The device was returned to the factory for evaluation on 07/20/2026.An investigation was conducted on 8/18/2026.A visual inspection was conducted.Signs of clinical use and no evidence of blood was observed.The delivery device and the intact seal and tension spring was returned outside the loading device with the white plunger not depressed and the blue safety lock on, which prevents the white plunger from being depressed.There were no visual defects observed on the intact seal or tension spring assembly.Measurements of the delivery device were taken; the inner diameter was measured at 0.195 inches; the outer diameter was measured at 0.219 inches (rm2036883).The length of the delivery tube was measured at 2.49 inches (mcv00004217).The measurement values recorded for the delivery tube were within the tolerance specifications.Based on the returned condition of the device as well as the evaluation results, the reported failures "fitting problem" and "detachment of device or device component" were not confirmed.The lot #3000500748 history record review was completed.There was an ncmr, rework, or deviation documented for the reported lot number.Based on the dhr/lhr review results, it was determined that there is a relation between the batch manufacturing process and the reported failures.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HST III System (3.8MM)
Common Device Name
Clamp, Vascular
Manufacturer (Section D)
Maquet Cardiovascular LLC
45 barbour pond drive
wayne NJ 07470
Manufacturer (Section G)
Maquet Cardiovascular LLC
45 barbour pond drive
wayne NJ 07470
Manufacturer Contact
arelean guzman
45 barbour pond drive
wayne, NJ 07470
MDR Report Key50137927
Report Number2242352-2026-0000654
Device Sequence Number10730486
Product Code DXC
UDI-Device Identifier00607567700314
UDI-Public(01)00607567700314(10)3000500748(17)260902
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K130382
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 08/05/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/02/2026
Device Model NumberHSK-3038
Device Catalogue NumberHSK-3038
Device Lot Number3000500748
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer07/20/2026
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/15/2026
Supplement Date Received by Manufacturer09/18/2026
Initial Report FDA Received Date08/05/2026
Supplement Report FDA Received Date09/21/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/02/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1