C.R. Bard, Inc. (BASD) -3006260740 PowerPort isp M.R.I. Implantable Port, ChronoFlex Single-Lumen, 8F; Port & Catheter, Implanted, Subcutaneous, Intravascular
|
Back to Search Results |
|
| Catalog Number |
1808060 |
| Medical Device Problem Codes |
Fracture (1260); Obstruction of Flow (2423)
|
| Health Effect - Clinical Codes |
Unspecified Infection (1930); Thrombosis/Thrombus (4440)
|
| Date of Event |
07/14/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported through the litigation process that, sometime after the port placement procedure, the catheter had allegedly fractured.It was further reported that, patient was diagnosed with an unspecified infection and blood clots.However, the current status of the patient was unknown.
|
| |
|
Event or Problem Description
|
|
It was reported through the litigation process that, sometime after the port placement procedure, the catheter had allegedly fractured.It was further reported that, patient was diagnosed with an unspecified infection and blood clots.However, the current status of the patient was unknown.
|
| |
|
Additional Manufacturer Narrative
|
|
H11: manufacturing review: the device history records have been reviewed, and this lot met all release criteria.Investigation summary: the physical device was not returned for evaluation.No medical records were provided for review.The investigation is inconclusive for the reported fracture, infection, blood clot issues as no objective evidence was provided for review.A review of sterilization records, endotoxin testing results, and bioburden data was conducted in accordance with applicable quality system and regulatory requirements, and no nonconformances or anomalies were identified.Based upon the available information, the definitive root cause is unknown.Labeling review: as the reported event did not allege a labeling or use related issue, a labeling review is not required
d4 (unique identifier (udi) #), g3, h6 (method)
section a through f: the information provide by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Additional Manufacturer Narrative
|
|
H11: as the lot number for the device was provided, a review of the device history records will be performed.The sample was not returned to the manufacturer for inspection/evaluation.Therefore, the investigation of the reported event is inconclusive.Based upon the available information, the definitive root cause for this event is unknown.The instructions for use (ifu) is adequate for the reported device/patient code(s) and provides general instructions for use, as well as warnings, precautions and potential complications associated with the device.Upon receipt of new or additional information, a follow-up report will be submitted as applicable.Section a through f : the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Search Alerts/Recalls
|
|
|