Maquet Critical Care AB BASE UNIT SERVO-U; Ventilator, Continuous, Facility Use Back to Search Results
Model Number 6694800
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/22/2026
Type of Reportable Event Malfunction
Event or Problem Description
Manufacturer's ref.#: (b)(4).
 
Additional Manufacturer Narrative
On-site investigation was performed by the field service engineer.According to received information, the replacement of the support arm is required to resolve the issue.The support arm serves to relieve the patient from the weight of the tubing system.According to the information received during the processing of the complaint, the arm was pulled to move the ventilator system.Therefore, the most probable cause of the issue is considered to be the user not following the instructions.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BASE UNIT SERVO-U
Common Device Name
Ventilator, Continuous, Facility Use
Manufacturer (Section D)
Maquet Critical Care AB
roentgenvagen 2
solna
Manufacturer (Section G)
Maquet Critical Care AB
roentgenvagen 2
solna
Manufacturer Contact
caroline kabbabe
roentgenvagen 2
solna 
MDR Report Key50138275
Report Number8010042-2026-0001064
Device Sequence Number18421782
Product Code CBK
UDI-Device Identifier07325710001110
UDI-Public(01)07325710001110(11)210208
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K201874
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Y
Reporter Type Manufacturer
Report Source Foreign,Distributor
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 08/05/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number6694800
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/14/2026
Supplement Date Received by Manufacturer09/21/2026
Initial Report FDA Received Date08/05/2026
Supplement Report FDA Received Date10/01/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured02/08/2021
Is the Device Labeled for Single Use? (Y/N) No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-