On (b)(6)2026, a patient prepared for an ultrasound guided percutaneous transhepatic cholangiography drainage (ptcd) catheter placement procedure using a navarre opti drain catheter.Prior to the procedure, it was noted that the sure-lok thread was allegedly broken.The procedure was completed using another device.There was no patient contact.
|
|
H11: manufacturing review: the device history records have been reviewed, and this lot met all release criteria.Investigation summary: although the sample was not returned for evaluation, two electronic photos were provided and reviewed.The first photo shows the partial view of the catheter hub.The thread was not noted within the catheter fiber hole.Second photo shows the partial view of catheter and one device packaging with label.The lot and product information were matched and verified with tw.Break cannot be verified as device shows the partial view of the device.The investigation is inconclusive for the reported catheter fiber break and separation issue as provided photo is not sufficient to confirm the issue.The definitive root cause could not be determined based upon available information.Labeling review: as the reported event did not allege a labeling or use related issue, a labeling review is not required.E1 (b)(6) university)
b5, d2, d4 (unique identifier (udi) #), g3, h6 (method, result, conclusion)
section a through f: the information provide by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|