It was reported that intra-op during the procedure, the rotation was not smooth, turning intermittently and unable to rotate fully into position, with a loud noise.There was no patient impact.
|
|
H3: product analysis found that the device, stylette, tyvek envelope, and an open pouch were returned in a resealable bag within a large plastic sleeve.Upon receipt, the pouch was found to be open on three of its four sides.No traces of contamination were observed on the returned device.However, the shaft between the inner and outer hubs was observed to be slightly bent.The inner assembly rotated freely by hand with no binding.The device was easily loaded into a handpiece, and wobbling was observed during the blade rotation in oscillating mode up to 5,000 rpm.The complaint was confirmed, due to an out of specification condition.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|