| Model Number |
Mi1200 SYNCHRONY |
| Medical Device Problem Code |
Impedance Problem (2950)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/29/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
The user's parent reported a sudden hearing loss with the right ci in (b)(6)2026.The user has been reimplanted.
|
| |
|
Event or Problem Description
|
|
The user's parent reported a sudden hearing loss with the right ci in (b)(6) 2026.The user has been reimplanted.
|
| |
|
Additional Manufacturer Narrative
|
|
The device has been explanted and should be returned to the manufacturer for evaluation.When available, a device failure analysis will be submitted as a follow up report.
|
| |
|
Additional Manufacturer Narrative
|
|
Conclusion: overload fractures in the active electrode which are consistent with an external mechanical impact were determined to be the root cause of device failure.The problems given in the recipient report appear to match the damage found.This is a final report.
|
| |
|
Search Alerts/Recalls
|