Model Number DVPB3D4
Medical Device Problem Codes Electromagnetic Interference (1194); Over-Sensing (1438)
Health Effect - Clinical Codes Chest Pain (1776); Paresthesia (4421)
Date of Event 07/17/2026
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the patient presented to the emergency room due to tingling in their arms, chest pain, and an alarm.The patient had been hosing off their patio at the time.An interrogation of the implantable cardioverter defibrillator (icd) indicated a lead integrity alert (lia) triggered due to high rate non-sustained episodes and short ventricular intervals.Review of the episodes indicated oversensing on the ventricular channel due to electromagnetic interference.The implantable cardioverter defibrillator (icd) remains in use. no further patient complications have been reported as a result of this event.
 
Additional Manufacturer Narrative
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
Cobalt¿ VR MRI SureScan¿
Common Device Name
IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (NON-CRT)
Manufacturer (Section D)
Medtronic Europe Sarl
route du molliau 31
case postale
tolochenaz vaud 1131
SZ  1131
Manufacturer (Section G)
Medtronic Europe Sarl
route du molliau 31
case postale
tolochenaz vaud 1131
SZ   1131
Manufacturer Contact
gerard torenvliet
8200 coral sea st ne
mounds view, MN 55112--439
+1(763)526-9277
MDR Report Key50140449
Report Number9614453-2026-03152
Device Sequence Number9408990
Product Code LWS
UDI-Device Identifier00763000178512
UDI-Public(01)00763000178512(17)250214(20)02
Combination Product (Y/N)N
Initial Reporter StateMO
Initial Reporter CountryUS
PMA/510(K) Number
P980016
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2024
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 08/05/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date02/14/2025
Device Model NumberDVPB3D4
Device Catalogue NumberDVPB3D4
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/17/2026
Initial Report FDA Received Date08/05/2026
Date Device Manufactured08/16/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
6935M55 :LEAD
Patient Age67 YR
Patient SexFemale
Patient Weight123 LB
-
-