• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION AMS 800 Artificial Urinary Sphincter; Device incontinence mechanical/hydraulic

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION AMS 800 Artificial Urinary Sphincter; Device incontinence mechanical/hydraulic Back to Search Results
Model Number 72400024
Medical Device Problem Codes Fluid/Blood Leak (1250); Material Puncture/Hole (1504)
Health Effect - Clinical Code Urinary Incontinence (4572)
Date of Event 07/07/2026
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the patient with an artificial urinary sphincter (aus) experienced recurrent urinary incontinence.A surgical procedure was performed, during which fluid loss was identified due to a hole in the balloon that was discovered when it was filled on the back table.The balloon and cuff were replaced.No additional patient complications were reported.
 
Event or Problem Description
It was reported that the patient with an artificial urinary sphincter (aus) experienced recurrent urinary incontinence.A surgical procedure was performed, during which fluid loss was identified due to a hole in the balloon that was discovered when it was filled on the back table.The balloon and cuff were replaced.No additional patient complications were reported.
 
Additional Manufacturer Narrative
Model number/catalog number: 72400024.Serial number: n/a.Batch/lot number: 627485009.Model/catalog description: balloon fgs 61-70 cm.Unique identifier (udi) #: (b)(4).Model number/catalog number: 72404127.Serial number: n/a.Batch/lot number: 627471010.Model/catalog description: control pump fgs iz.Unique identifier (udi) #: (b)(4).
 
Additional Manufacturer Narrative
This report is being submitted to correct the concomitant information in additional mfr narrative field (h11) in section h, device manufactures only.Model number/catalog number: 72404131.Serial number: n/a.Batch/lot number: 625223011.Model/catalog description: belt cuff fgs 4.5 cm iz.Unique identifier (udi) #: (b)(4).Model number/catalog number: 72404127.Serial number: n/a.Batch/lot number: 627471010.Model/catalog description: control pump fgs iz.Unique identifier (udi) #: (b)(4).The device is not available for analysis; therefore, no physical analysis of the product could be performed.The reported device performance allegation cannot be confirmed.The device history record (dhr) review confirmed that the device met all material, assembly and performance specifications.A review of the device instructions for use (ifu) was completed and did not reveal any evidence of device misuse, off label use, or failure to follow instructions.The reported patient symptom of urinary incontinence is known risk associated with this device type and indicated as such in the instructions for use.A risk review was completed and confirmed that the event of urinary incontinence, hole/ perforated, and fluid leak was defined in the product risk documentation.This event type has been accounted for during analysis to support acceptable risk benefit for the product.Based on the information available, a conclusion code of end of life problem identified was assigned to this investigation, because the reason for the fluid leak was most likely determined to be the artificial urinary sphincter (aus) device reaching end of its useful life from the analysis of the available information.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AMS 800 Artificial Urinary Sphincter
Common Device Name
Device incontinence mechanical/hydraulic
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
Boston Scientific Corporation
300 boston scientific way
marlborough MA 01752
Manufacturer Contact
farshad fahimi
4100 hamline avenue north
building c
saint paul, MN 55112
MDR Report Key50141208
Report Number2124215-2026-41878
Device Sequence Number9350006
Product Code EZY
UDI-Device Identifier00878953003092
Combination Product (Y/N)Y
Initial Reporter StateMD
Initial Reporter CountryUS
PMA/510(K) Number
P000053
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2010
Device Explanted Year2026
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 09/17/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Expiration Date12/04/2014
Device Model Number72400024
Device Catalogue Number72400024
Device Lot Number627485009
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/07/2026
Supplement Date Received by Manufacturer08/25/2026
Initial Report FDA Received Date08/05/2026
Supplement Report FDA Received Date09/17/2026
Date Device Manufactured12/18/2009
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age85 YR
Patient SexMale
-
-