| Model Number |
480522-02 |
| Medical Device Problem Codes |
Thermal Decomposition of Device (1071); Therapy Delivered to Incorrect Body Area (1508)
|
| Health Effect - Clinical Code |
Burn(s) (1757)
|
| Date of Event |
07/08/2026
|
|
Type of Reportable Event
|
Other
|
|
Event or Problem Description
|
|
It was reported that during a da vinci-assisted benign hysterectomy surgical procedure, the vessel sealer curved instrument was hot and blanched the urachus while sealing the round ligament.The surgeon did not perform any interventions and reported it as minor non-consequential thermal spread.The procedure was completed as planned.After the procedure, cystoscopy was performed to assess ureteral jets and they were fine.
|
| |
|
Additional Manufacturer Narrative
|
|
The vessel sealer curved instrument used during this event was not received for failure analysis investigations.The vessel sealer curved instrument lot number: k11260528-0198 was installed 1 time on universal surgical manipulator (usm) 2.E-200 generator logs showed a total of 168 seal events with 2 high z errors, indicating a high impedance situation either due to excessive bio-debris between the jaws or excessive tissue during sealing.The instrument was used for about 26 minutes.The system logs don¿t show any instrument related errors.
|
| |
|