Product event summary: the afr-00001 catheter with lot number 0013331049 was returned and analyzed.No anomalies were identified during external visual inspection of the catheter.The catheter passed the mapping and ablation test with the mapping system (test/lab), no errors were triggered.No performance issues were identified with the catheter.During deflection testing, pushing and pulling steering wings was possible without any anomalies.Leak test was performed to check for any leaks on the irrigation line.The pressure decay was 0.015 pounds per square inch (psi), the catheter passed the test.Occlusion test was performed to check for any blocks in the irrigation line.The pressure drop was 23.912, the catheter passed the test.The catheter passed the shorts test.No errors were triggered.The catheter passed the mapping test.No errors were triggered.No anomalies were identified during the inspection of the sphere segment.All the electrode wires were intact.In conclusion, the reported issues (steam pop, temp rise) were not repr oduced during testing or inspection.The catheter passed the returned product inspection.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|