On (b)(6) 2026, a patient underwent a thrombectomy and atherectomy procedure using a rotarex device for the treatment of superficial femoral artery (sfa) and popliteal artery disease.During the procedure, it was reported that the guidewire fractured was noticed to be broken off, resulting in retention of a wire fragment within the patient.A covered stent was subsequently implanted to entrap the retained fragment.The patient's current condition was reported as good.
|
|
H11: manufacturing review: a manufacturing review was conducted and there was nothing found to indicate there was a manufacturing related cause for this event.Investigation summary: no physical sample / images / videos were received.A physical investigation was not possible.Due to no sample / images / videos received, the reported malfunction cannot be confirmed.A possible reason for the breakage of the guide wire could be insufficient drainage flow through the catheter which leads to heat development inside the catheter, or as it is stated in the report abnormal rapid move of the patient.The guidewire and helix tend to break easily when set under stress.A clear root cause could not be identified but a damaged guidewire represents a known inherent risk.Labeling review: as the reported event did not allege a labelling or use related issue, a labelling review is not required.D2b (mcw; dqx), e1, g3, h6 (component).Section a through f: the information provide by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|