Terumo medical corporation received an fda medwatch report # 2429304-2026-00102.The event description states: " it was reported a patient was scheduled for an endoscopic retrograde cholangiopancreatography (ercp) procedure due to elevated bilirubin and being jaundiced.The ercp was performed under general anesthesia and was unsuccessful because the submucosal injection had failed and the ampulla could not be accessed.The procedure was then converted to an endoscopic ultrasonography (eus) rendezvous.During the eus rendezvous, the dilated duct was accessed with a 19-gauge eus needle.The angled single use guidewire was fed through the needle in an attempt to gain access to the bile duct.The wire was pushed into the lumen of the small bowel.At that point, the tip of the wire broke off.The case was aborted.The broken piece of the wire was left inside the patient.The patient was an inpatient and remained an inpatient after the aborted procedure.An ercp procedure would be repeated in a few days to allow for inflammation from the submucosal injection to be reduced.Additional information about the location of the broken guidewire tip was requested.Conflicting information was received.One report indicated the tip of the guidewire had broken off into the liver, while another report indicated the tip of the guidewire had broken off in the duodenum and the physician expected it to pass naturally.Clarification has been requested along with additional details on the patient outcome and details of the additional ercp procedure.No further details were provided at the time of this report"
terumo medical corporation received a supplemental fda medwatch report # 2429304-2026-00102 on 31jul2026.The event description states: "additional information was received from the customer.It was reported the tip of the wire broke off in the pancreatic duct (pd).The pd was accessed with a 19-gauge needle, and an attempt was made to advance the wire antegrade to the duodenum trans-papillary.The broken piece of the wire was left in the main pd as the physician though it would pass naturally through the gastrointestinal (gi) system.There was no harm caused to the patient due to the broken wire.The procedure was not aborted, but rather it was unsuccessful in traversing the pd stricture both antegrade and retrograde.The procedure was performed under general anesthesia.The procedure was not repeated.The indication for the procedure was chronic pancreatitis, pd stones, and pd stricture.The broken wire did not result in any prolonged hospitalization for the patient.The patient's current condition was stable.This supplemental report is being submitted to provide additional information from the customer.Updated fields: b5, f7, and fl0.Corrections to bl and b2.This event was initially reported as a serious injury; however, upon further follow-up with the customer, it was determined the device did not contribute to a serious injury as there was no harm to the patient.Bl should not be marked as an adverse event, and b2 should not be marked at all.".
|