(b)(4)
date sent 10/6/2026
d4 batch # 905d08
investigation summary
the product was returned for thorough evaluation, which included visual inspections and functional testing of the returned device.Visual analysis of the returned sample determined that the cdh31p device arrived with no apparent damage along with the anvil and stapler retainer inside a plastic bag.The breakaway washer was present, uncut and the device was fully loaded with staples, indicating that the device had not been fired.No battery was received, however, when the test battery was installed the green checkmark illuminated indicated that the device was not reset.No functional test was performed due to the condition of the device.The device was disassembled to verify the condition of the internal components and the actuator was touching the stop switch actuator.In addition, a tyvek returned along with the device.This defect has been potential to a manufacturing process.Based on the information currently available, a product issue was identified during the investigation of the sample received.This product issue will be addressed through ethicon endo surgery¿s quality system.A manufacturing record evaluation was performed for the finished device batch number 905d08, and no non-conformances related to the reported complaint condition were identified.
|