• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. Bard, Inc. (BASD) -3006260740 PowerPort isp M.R.I. Implantable Port, Groshong Single-Lumen, 8F; Port & Catheter, Implanted, Subcutaneous, Intravascular

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. Bard, Inc. (BASD) -3006260740 PowerPort isp M.R.I. Implantable Port, Groshong Single-Lumen, 8F; Port & Catheter, Implanted, Subcutaneous, Intravascular Back to Search Results
Catalog Number 9808560
Medical Device Problem Codes Difficult to Flush (1251); Device Appears to Trigger Rejection (1524); Improper or Incorrect Procedure or Method (2017); Obstruction of Flow (2423)
Health Effect - Clinical Code Thrombosis/Thrombus (4440)
Date of Event 06/29/2026
Type of Reportable Event Serious Injury
Event or Problem Description
On (b)(6) 2026, a patient underwent a port placement procedure using a powerport m.R.I.Isp.On (b)(6) 2026, during an outpatient visit, the port was accessed; however, medication could not be infused through the device.The port was subsequently removed, and a blood like substance was observed adherent near the valve slit at the catheter tip.Following transection of the catheter tip, patency was restored and normal flow was confirmed by flushing the port with normal saline.The obstruction was determined to be due to a clog at the catheter tip.It was reported that blood resembling a clot was adhering to the slit in the catheter valve at the time of removal.The device was retrieved for evaluation.The current status of the patient is unknown.
 
Additional Manufacturer Narrative
H11: as the lot number for the device was not provided, a review of the device history records could not be performed.The device has been returned to the manufacturer for evaluation.The investigation of the reported event is currently underway.Section a through f: the information provide by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PowerPort isp M.R.I. Implantable Port, Groshong Single-Lumen, 8F
Common Device Name
Port & Catheter, Implanted, Subcutaneous, Intravascular
Manufacturer (Section D)
C.R. Bard, Inc. (BASD) -3006260740
605 north 5600 west
salt lake city 84116
Manufacturer (Section G)
Bard Reynosa S.A. de C.V. -9617592
blvd. montebello #1
parque industrial colonial
reynosa, tamaulipas 88780
MX   88780
Manufacturer Contact
michael gaffney
800 w. rio salado pkwy
tempe, AZ 85281
+1(480)303-2689
MDR Report Key50144335
Report Number3006260740-2026-05178
Device Sequence Number13444359
Product Code LJT
UDI-Device Identifier00801741027512
UDI-Public(01)00801741027512
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K063377
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2026
Reporter Type Manufacturer
Report Source Other,Foreign,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 08/24/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue Number9808560
Device Lot NumberUnknown
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer07/13/2026
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/29/2026
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date08/05/2026
Supplement Report FDA Received Date09/04/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-