Beckman Coulter, Inc. AU5800 Chemistry Analyzer; Analyzer, Chemistry (Photometric, Discrete), For Clinical Use
|
Back to Search Results |
|
| Model Number |
AU5800-20 CHEMISTRY ANALYZER AU5800 Automation Ready |
| Medical Device Problem Code |
High Test Results (2457)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/16/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The customer reported the generation of erroneous high glucose (glu) results for twenty-nine (29) patients on their au5800 clinical chemistry analyzer.The samples were retested on another au5800 analyzer with results considered correct.There was no report of change to treatment, injury, or death associated with event.The customer did not provide patient demographics.The customer provided data for 29 patients,.
|
| |
|
Additional Manufacturer Narrative
|
|
Beckman coulter field service engineer (fse) evaluated the au5800 chemistry analyzer and identified the probable cause as a bent sample probe.The fse replaced the sample probe to resolve the issue.Section a2, a3, a4, a5 and a6: information not provided by customer.The beckman coulter internal identifier is (b)(4).
|
| |
|