BOSTON SCIENTIFIC CORPORATION VersaCross Connect LAAC Access Solution; Catheter, Septostomy
|
Back to Search Results |
|
| Model Number |
VXAK0107 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Codes |
Low Blood Pressure/ Hypotension (1914); Cardiac Perforation (2513); Pericardial Effusion (3271)
|
| Date of Event |
07/13/2026
|
|
Type of Reportable Event
|
Death
|
|
Event or Problem Description
|
|
The patient died.It was reported that during a watchman procedure left atrial appendage closure (laac).The doctor was waiting until he got the fluro shot of the appendage to see which rep/device he was going to use.The patient at the beginning of the procedure had a baseline effusion, and very calcified valves.He went transseptal with a versacross fixed system, and then lost access, resulting in him going back to the right side.He asked for a trusteer due to the thickness of the septum, and right as he put it up, he noticed a drop in pressure.The patient had suffered a pericardial effusion, and the doctor immediately grabbed a pericardiocentisis kit , also open heart surgery and auto-transfusion were performed.However, the patient's act was 540, and after 20 minutes and a lot of protamine, it still was at 477.The patient was then wheeled off to the or for ct surgery.It was learned that the patient had passed.It is unknown the exact cause of death, but the doctor did not credit it to any device/sheath.
|
| |
|
Event or Problem Description
|
|
The patient died.It was reported that during a watchman procedure left atrial appendage closure (laac).The doctor was waiting until he got the fluro shot of the appendage to see which rep/device he was going to use.The patient at the beginning of the procedure had a baseline effusion, and very calcified valves.He went transseptal with a versacross fixed system, and then lost access, resulting in him going back to the right side.He asked for a trusteer due to the thickness of the septum, and right as he put it up, he noticed a drop in pressure.The patient had suffered a pericardial effusion, and the doctor immediately grabbed a pericardiocentesis kit, also open-heart surgery and auto-transfusion were performed.However, the patient's act was 540, and after 20 minutes and a lot of protamine, it still was at 477.The patient was then wheeled off to the or for ct surgery.It was learned that the patient had passed.It is unknown the exact cause of death, but the doctor did not credit it to any device/sheath.
|
| |
|
Additional Manufacturer Narrative
|
|
We are unable to provide the complete unique identifier (udi) at the time of the submission of this report, as it was not yet available.If additional details become available, a supplemental report will be submitted.
|
| |
|
Event or Problem Description
|
|
The patient died.It was reported that during a watchman procedure left atrial appendage closure (laac).The doctor was waiting until he got the fluro shot of the appendage to see which rep/device he was going to use.The patient at the beginning of the procedure had a baseline effusion, and very calcified valves.He went transseptal with a versacross fixed system, and then lost access, resulting in him going back to the right side.He asked for a trusteer due to the thickness of the septum, and right as he put it up, he noticed a drop in pressure.The patient had suffered a pericardial effusion, and the doctor immediately grabbed a pericardiocentesis kit, also open-heart surgery and auto-transfusion were performed.However, the patient's act was 540, and after 20 minutes and a lot of protamine, it still was at 477.The patient was then wheeled off to the operating room for ct surgery.It was learned that the patient had passed.It is unknown the exact cause of death, but the doctor did not credit it to any device/sheath.Device was disposed.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow up report is being submitted with an update to sections:
a1, a2, a3a and a3b: patient information.B1: adverse event/product problem
d1: brand name.D3: manufacturer information.D4: lot number.G1: manufacturer site information.G4: premarket / 510(k).H4: device manufacture date.H6: impact codes.
|
| |
|
Additional Manufacturer Narrative
|
|
Correction to section h6 (evaluation conclusion codes).Additional information to section b5.Investigation summary:
with all the available information, boston scientific is unable to confirm cause of the reported clinical observation of pericardial effusion, hypotension, perforation, cardiac and death.The device has not been received for analysis; therefore, a technical analysis of the complaint device could not be completed.Device history record review:
it was confirmed this device met specification prior to distribution and there were no manufacturing deviations which could have contributed to the reported event.Risk review:
a risk review performed for the versacross connect device confirmed that the event of pericardial effusion, hypotension, perforation, cardiac and death are known events defined in the product risk management documentation.This event type has been accounted for during product risk analysis to support acceptable risk benefit for the product.Labeling review:
based on the information provided, there is no evidence that the device was used in a manner inconsistent with the labelled indications/instructions for use (ifu).Investigation conclusion:
boston scientific has assigned an investigation conclusion code of no problem detected and known inherent risk of device and supporting evidence that came to this conclusion.Boston scientific investigation findings were unable to establish a clear conclusion about the cause of the reported event.The device met all requirements prior to being approved for final distribution/sale and there was no evidence to suggest that the instructions for use (ifu) was not followed.
|
| |
|