| Model Number |
TSGC0801 |
| Medical Device Problem Code |
Air/Gas in Device (4062)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/14/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that on (b)(6) 2026, a patient presented with functional tricuspid regurgitation (tr) for a triclip procedure.During device preparation, bubbles were identified when the triclip was inserted within the clip introduction.Troubleshooting was performed and the clip was readjusted, during which the clip became caught on the clip introducer and tore the device with the grippers.The tcds was removed from the patient and a replacement triclip was selected and successfully implanted.Two additional triclips were implanted successfully.The procedure was discontinued, the final tr was reduced.There were no adverse patient effects or clinically significant delays reported.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was not returned for analysis.A review of the lot history record revealed no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history did not indicate product quality issue.Based on the available information, a cause for the reported air could not be determined.There is no indication of a product quality issue with respect to manufacture, design, or labeling.
|
| |
|