| Model Number |
HexaVue IP |
| Medical Device Problem Code |
Display or Visual Feedback Problem (1184)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/07/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
The user facility reported ¿no video from input¿ with the monitor to their hexavue ip custom room rack during a patient procedure, resulting in procedure delay.No report of injury.
|
| |
|
Additional Manufacturer Narrative
|
|
A steris service technician arrived onsite to inspect the hexavue ip custom room rack and found the lmm/mdm compositor needed to be configured.To resolve the issue the technician configured the lmm/mdm compositor, tested the function and operation of the unit, confirmed it to be operating to specification, and returned the hexavue ip custom room rack to service.No additional issues have been reported.
|
| |
|