Abbott Medical (CRM-Sylmar) DURATA STS OPTIM ACTIVE FIXATION; Implantable Cardioverter Defibrillator (Non-Crt) Back to Search Results
Model Number 7122/65
Medical Device Problem Codes Failure to Capture (1081); Device Sensing Problem (2917)
Health Effect - Clinical Codes Cardiac Perforation (2513); Pericardial Effusion (3271)
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that during remote follow-up, the patient's right ventricular (rv) lead exhibited a low r-wave amplitude.During in-clinic testing, loss of capture was also noted on the rv lead.X-ray and computed tomography (ct) imaging revealed that the rv lead had perforated the myocardium and pericardium, with the tip of the lead located in the left pericardial fat.The physician elected to explant the rv lead and replace it with a new one.Following the procedure, a pericardial effusion was observed, and the patient underwent a sternotomy with attempted repair.The patient condition remained stable.
 
Additional Manufacturer Narrative
A device history record (dhr) review was performed and all required manufacturing processes and inspections steps were confirmed to be completed per the requirements.The device met specifications prior to leaving abbott manufacturing facilities.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DURATA STS OPTIM ACTIVE FIXATION
Common Device Name
Implantable Cardioverter Defibrillator (Non-Crt)
Manufacturer (Section D)
Abbott Medical (CRM-Sylmar)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
Abbott Medical (CRM-Sylmar)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
navpreet singh
15900 valley view court
sylmar, CA 91342
MDR Report Key50148127
Report Number2017865-2026-16285
Device Sequence Number18493631
Product Code LWS
UDI-Device Identifier05414734502153
UDI-Public(01)05414734502153(10)A000179853(17)280930
Combination Product (Y/N)Y
Initial Reporter CountryCA
PMA/510(K) Number
P950022
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2026
Device Explanted Year2026
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 08/05/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number7122/65
Device Lot NumberA000179853
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/14/2026
Initial Report FDA Received Date08/05/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured10/24/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention; Life Threatening;
Patient SexUnknown
-
-