| Model Number |
0185 |
| Medical Device Problem Code |
High impedance (1291)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/21/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that a review of the right ventricular (rv) lead showed evidence of a gradual increase in shock impedance that remains within normal limits.Technical services provided troubleshooting recommendations, and reprogramming options were discussed.It was decided that the lead would be surgically abandoned and replaced.Additionally, a segment of the lead was returned for analysis.No additional adverse patient effects were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
This product investigation is still in progress.This report will be updated when product investigation is complete.
|
| |
|