Model Number D120
Medical Device Problem Codes Pacing Problem (1439); Inappropriate/Inadequate Shock/Stimulation (1574); Data Problem (3196)
Health Effect - Clinical Code Shock from Patient Lead(s) (3162)
Date of Event 09/28/2024
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that this implantable cardioverter defibrillator (icd) delivered inappropriate anti-tachycardia pacing (atp) during a sinus tachycardia episode.Upon further review of the episode, boston scientific technical services (ts) discussed the rhythmmatch was not appropriately correlated due to the misalignment of the template with the shock egm.Reprogramming options were discussed.No adverse patient effects were reported.This device remains in service.The complaint device was not returned to boston scientific, therefore, product analysis could not be performed.Additional information was received that this device was suspected of delivering inappropriate atp and electric shocks.Further evaluation by cardiology was recommended.No additional adverse patient effects were reported.This device remains in service.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MOMENTUM? EL ICD VR
Common Device Name
Implantable cardioverter defibrillator (ICD)
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
Boston Scientific Corporation
cashel road
clonmel E91 T 862
EI   E91 T862
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
+1(651)582-6168
MDR Report Key50150248
Report Number2124215-2026-42070
Device Sequence Number13412472
Product Code LWS
UDI-Device Identifier00802526587702
UDI-Public(01)00802526587702(17)250424
Combination Product (Y/N)N
Initial Reporter StateNC
Initial Reporter CountryUS
PMA/510(K) Number
P960040
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2023
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 08/26/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date04/24/2025
Device Model NumberD120
Device Catalogue NumberD120
Device Lot Number217390
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/15/2026
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date08/05/2026
Supplement Report FDA Received Date09/04/2026
Date Device Manufactured05/04/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age45 YR
Patient SexFemale
-
-