It was reported that this implantable cardioverter defibrillator (icd) delivered inappropriate anti-tachycardia pacing (atp) during a sinus tachycardia episode.Upon further review of the episode, boston scientific technical services (ts) discussed the rhythmmatch was not appropriately correlated due to the misalignment of the template with the shock egm.Reprogramming options were discussed.No adverse patient effects were reported.This device remains in service.The complaint device was not returned to boston scientific, therefore, product analysis could not be performed.Additional information was received that this device was suspected of delivering inappropriate atp and electric shocks.Further evaluation by cardiology was recommended.No additional adverse patient effects were reported.This device remains in service.
|