MAUDE Adverse Event Report: Maquet Cardiovascular LLC HST III System (3.8MM); Clamp, Vascular

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Maquet Cardiovascular LLC HST III System (3.8MM); Clamp, Vascular Back to Search Results
Model Number HSK-3038
Medical Device Problem Code Fitting Problem (2183)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/15/2026
Type of Reportable Event Malfunction
Event or Problem Description
N/a.
 
Additional Manufacturer Narrative
Tw (b)(4) updated sections: b4, g3, g6, h2, h3, h6, h11.The device was returned to the factory for evaluation on 07/20/2026.Photographs were provided.Product information was observed.The seal was observed in the loading device body.The delivery device was observed outside the loading device.An investigation was conducted on 8/19/2026.A visual inspection was conducted.Signs of clinical use and no evidence of blood was observed.The delivery device was returned inside the loading device with the white plunger depressed and the blue safety lock off, which allows for the white plunger to be depressed.The delivery device was removed from the loading device.The seal and tension spring assembly remained inside the loading device.The seal and tension spring assembly were removed from the loading device with no physical or visual difficulties observed.The seal was observed to be unraveled at the center of the seal.The seal was also observed to be detached from the tether with the blue string still attached.Measurements of the delivery device were taken; the inner diameter was measured at 0.195 inches, the outer diameter was measured at 0.219 inches (rm2036883).The length of the delivery tube was measured at 2.49 inches (mcv00004217).The measurement values recorded for the delivery tube were within the tolerance specifications.Based on the returned condition of the device as well as the evaluation results and the photographic evaluation, the reported failure "fitting problem" was confirmed as well as the analyzed failures "premature deployment", "detachment of device or device component" and "unraveled seal" were observed.The lot # 3000541610 history record review was completed.There were no ncmrs, rework, or deviations documented for the reported lot number.Based on the dhr/lhr review results, it was determined that there is no relation between the batch manufacturing process and the reported failure or the analyzed failures.Specific actions for the reported failure mode are being maintained and documented under maquet's corrective and preventive action (capa) system.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HST III System (3.8MM)
Common Device Name
Clamp, Vascular
Manufacturer (Section D)
Maquet Cardiovascular LLC
45 barbour pond drive
wayne NJ 07470
Manufacturer (Section G)
Maquet Cardiovascular LLC
45 barbour pond drive
wayne NJ 07470
Manufacturer Contact
arelean guzman
45 barbour pond drive
wayne, NJ 07470
MDR Report Key50151644
Report Number2242352-2026-0000661
Device Sequence Number9407383
Product Code DXC
UDI-Device Identifier00607567700314
UDI-Public(01)00607567700314(10)3000541610(17)270305
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K130382
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 08/06/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberHSK-3038
Device Catalogue NumberHSK-3038
Device Lot Number3000541610
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer07/20/2026
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/16/2026
Supplement Date Received by Manufacturer09/10/2026
Initial Report FDA Received Date08/06/2026
Supplement Report FDA Received Date09/17/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/05/2026
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-