• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: Maquet Cardiovascular LLC Acrobat-i Stabilizer; Stabilizer, Heart

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Maquet Cardiovascular LLC Acrobat-i Stabilizer; Stabilizer, Heart Back to Search Results
Model Number OM-10000Z
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/14/2026
Type of Reportable Event Malfunction
Event or Problem Description
N/a.
 
Additional Manufacturer Narrative
Tw # (b)(4).Updated sections: b4,d9,e1-email,g3,g6,h2,h3,h6,h11 the device was returned to the factory for evaluation on 09/15/2026.An investigation was conducted on 09/17/2026.A visual inspection was conducted.Signs of clinical use and no evidence of blood was observed.The top part of the mount was observed to be be broken off from the mount.No other visual defects were observed.The knob was evaluated for its ability to tighten the flexlink arm while the locking lever was in both the locked and unlocked positions.The flexlink arm was able to be secured in position when the knob was turned clockwise.The knob did tightened the arm.Based on the returned condition of the device as well as the evaluation results, the reported failure "break" was confirmed.The lot #3000514015 history record review was completed.There were no ncmrs, rework, or deviations documented for the reported lot number.Based on the dhr/lhr review results, it was determined that there is no relation between the batch manufacturing process and the reported failure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
Acrobat-i Stabilizer
Common Device Name
Stabilizer, Heart
Manufacturer (Section D)
Maquet Cardiovascular LLC
45 barbour pond drive
wayne NJ 07470
Manufacturer (Section G)
Maquet Cardiovascular LLC
45 barbour pond drive
wayne NJ 07470
Manufacturer Contact
arelean guzman
45 barbour pond drive
wayne, NJ 07470
MDR Report Key50152612
Report Number2242352-2026-0000662
Device Sequence Number9341636
Product Code MWS
UDI-Device Identifier00607567100008
UDI-Public(01)00607567100008(10)3000514015(17)271103
Combination Product (Y/N)N
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 08/06/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberOM-10000Z
Device Catalogue NumberOM-10000Z
Device Lot Number3000514015
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/15/2026
Supplement Date Received by Manufacturer08/06/2026
Initial Report FDA Received Date08/06/2026
Supplement Report FDA Received Date09/02/2026
Date Device Manufactured11/03/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Patient Age57 YR
Patient SexFemale
Patient Weight160 LB
-
-