Tw # (b)(4).Updated sections: b4,d9,e1-email,g3,g6,h2,h3,h6,h11
the device was returned to the factory for evaluation on 09/15/2026.An investigation was conducted on 09/17/2026.A visual inspection was conducted.Signs of clinical use and no evidence of blood was observed.The top part of the mount was observed to be be broken off from the mount.No other visual defects were observed.The knob was evaluated for its ability to tighten the flexlink arm while the locking lever was in both the locked and unlocked positions.The flexlink arm was able to be secured in position when the knob was turned clockwise.The knob did tightened the arm.Based on the returned condition of the device as well as the evaluation results, the reported failure "break" was confirmed.The lot #3000514015 history record review was completed.There were no ncmrs, rework, or deviations documented for the reported lot number.Based on the dhr/lhr review results, it was determined that there is no relation between the batch manufacturing process and the reported failure.
|