• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: INTUITIVE SURGICAL, INC da Vinci; TIP COVER ACCESSORY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

INTUITIVE SURGICAL, INC da Vinci; TIP COVER ACCESSORY Back to Search Results
Model Number 400180-16
Medical Device Problem Code Material Separation (1562)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/09/2026
Type of Reportable Event Serious Injury
Event or Problem Description
Refer to h11 for follow-up information.
 
Event or Problem Description
It was reported that during a da vinci¿assisted partial nephrectomy, the tip cover accessory from the monopolar curved scissors (mcs) was reported to have dislocated into the abdominal cavity.The tip cover accessory was retrieved during the same procedure using laparoscopic forceps.No patient injury was reported.It is unknown whether the mcs instrument collided with other instruments.The tip cover accessory appeared to have been properly installed, with no orange surface visible after installation, and no electrolube or other lubricant was applied prior to installation.Any resistance or difficulty during instrument removal was unknown, and it was also unknown whether any damage to the tip cover accessory, mcs instrument, or cannula was noted after the event.No additional surgical procedure was required to remove the fragment, and no post-operative imaging (e.G., x-ray or ultrasound) was performed to assess for retained fragments.The patient had not returned due to any post-surgical complications related to the event.
 
Additional Manufacturer Narrative
Device evaluation: intuitive surgical, inc.(isi) received the monopolar curved scissor (mcs) tip cover accessory to perform failure analysis.Using an installation tool, the mcs tip cover accessory was placed on an mcs instrument.The mcs tip cover accessory was loosely attached to the scissors and could be removed easily with little effort.An additional observation not reported by customer was that the mcs tip cover accessory was found to have tearing 25-31 mm from the proximal end where the instrument inserts.The tears were axially aligned with the tip cover accessory and measured from 1mm to 3mm in length.There were no signs of thermal damage present at the end of any of the tears.
 
Additional Manufacturer Narrative
The monopolar curved scissors (mcs) tip cover accessory has not been received for failure analysis evaluation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
da Vinci
Common Device Name
TIP COVER ACCESSORY
Manufacturer (Section D)
INTUITIVE SURGICAL, INC
3410 central expressway
santa clara CA 95051
Manufacturer (Section G)
INTUITIVE SURGICAL, INC
3410 central expressway
santa clara CA 95051
Manufacturer Contact
izabel nielson
3410 central expressway
santa clara, CA 95051
+1(408)523-2100
MDR Report Key50155739
Report Number2955842-2026-34793
Device Sequence Number9103415
Product Code NAY
UDI-Device Identifier10886874111045
UDI-Public(01)10886874111045(10)K15250403
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K112263
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign
Initial Reporter Occupation Non-Healthcare Professional
Remedial Action Other
Type of Report Initial,Followup
Report Date (Section B) 09/23/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Operator of Device Health Professional
Device Model Number400180-16
Device Catalogue Number400180
Device Lot NumberK15250403
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer08/09/2026
Is the Reporter a Health Professional? No
Initial Date Received by Manufacturer 07/13/2026
Supplement Date Received by Manufacturer08/31/2026
Initial Report FDA Received Date08/06/2026
Supplement Report FDA Received Date09/23/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/03/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
da Vinci Instruments and Accessories.
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-