| Model Number |
400180-16 |
| Medical Device Problem Code |
Material Separation (1562)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/09/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
Refer to h11 for follow-up information.
|
| |
|
Event or Problem Description
|
|
It was reported that during a da vinci¿assisted partial nephrectomy, the tip cover accessory from the monopolar curved scissors (mcs) was reported to have dislocated into the abdominal cavity.The tip cover accessory was retrieved during the same procedure using laparoscopic forceps.No patient injury was reported.It is unknown whether the mcs instrument collided with other instruments.The tip cover accessory appeared to have been properly installed, with no orange surface visible after installation, and no electrolube or other lubricant was applied prior to installation.Any resistance or difficulty during instrument removal was unknown, and it was also unknown whether any damage to the tip cover accessory, mcs instrument, or cannula was noted after the event.No additional surgical procedure was required to remove the fragment, and no post-operative imaging (e.G., x-ray or ultrasound) was performed to assess for retained fragments.The patient had not returned due to any post-surgical complications related to the event.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: intuitive surgical, inc.(isi) received the monopolar curved scissor (mcs) tip cover accessory to perform failure analysis.Using an installation tool, the mcs tip cover accessory was placed on an mcs instrument.The mcs tip cover accessory was loosely attached to the scissors and could be removed easily with little effort.An additional observation not reported by customer was that the mcs tip cover accessory was found to have tearing 25-31 mm from the proximal end where the instrument inserts.The tears were axially aligned with the tip cover accessory and measured from 1mm to 3mm in length.There were no signs of thermal damage present at the end of any of the tears.
|
| |
|
Additional Manufacturer Narrative
|
|
The monopolar curved scissors (mcs) tip cover accessory has not been received for failure analysis evaluation.
|
| |
|