• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: Medtronic Puerto Rico Villalba SelectSecure MRI SureScan; DRUG ELUTING PERMANENT RIGHT VENTRICULAR (RV)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Medtronic Puerto Rico Villalba SelectSecure MRI SureScan; DRUG ELUTING PERMANENT RIGHT VENTRICULAR (RV) Back to Search Results
Model Number 383074
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Wound Dehiscence (1154); Erythema (1840); Pain (1994); Pocket Erosion (2013); Discomfort (2330); Swelling/ Edema (4577); Medical device site infection (4845)
Date of Event 07/01/2026
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that that approximately two weeks post-implant the patient's pocket wound separated.The patient experienced a pocket infection, pocket erosion, discomfort, pain, redness, swelling, and wound dehiscence.The implantable pulse generator (ipg) system was explanted and replaced. no further patient complications have been reported as a result of this event.
 
Additional Manufacturer Narrative
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SelectSecure MRI SureScan
Common Device Name
DRUG ELUTING PERMANENT RIGHT VENTRICULAR (RV)
Manufacturer (Section D)
Medtronic Puerto Rico Villalba
road 149 km 56.3
villalba PR 00766
Manufacturer (Section G)
Medtronic Puerto Rico Villalba
road 149 km 56.3
villalba PR 00766
Manufacturer Contact
gerard torenvliet
8200 coral sea st ne
mounds view, MN 55112--439
+1(763)526-9277
MDR Report Key50155896
Report Number2649622-2026-23180
Device Sequence Number21151772
Product Code NVN
UDI-Device Identifier00763000740276
UDI-Public(01)00763000740276(17)280305(20)01
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
P030036
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2026
Device Explanted Year2026
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Type of Report Initial
Report Date (Section B) 08/06/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Model Number383074
Device Catalogue Number383074
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/20/2026
Initial Report FDA Received Date08/06/2026
Date Device Manufactured03/17/2026
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
5076-58 LEAD, W1DR01 IPG
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age49 YR
Patient SexMale
-
-