| Model Number |
HexaVue IP |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/09/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The user facility reported that during a patient procedure an employee inadvertently bumped into their hexavue ip custom room rack which caused the rack to disconnect.When the rack was plugged back in, the system ¿locked up and appeared to shut down.¿ the procedure was completed successfully following a short delay.No report of injury.
|
| |
|
Additional Manufacturer Narrative
|
|
A steris service technician arrived onsite following the reported event and rebooted the rack.The unit was tested, confirmed to be operating according to specification, and returned to service.No additional issues have been reported.
|
| |
|