Model Number CNS-2101
Medical Device Problem Codes Device Alarm System (1012); Protective Measures Problem (3015)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/13/2026
Type of Reportable Event Serious Injury
Event or Problem Description
The biomedical engineer (bme) reported that the central nurse station (cns) did not alarm that a patient's spo2 readings were low and the patient coded, but they did not expire.The end users stated they did not silence or turn off the spo2 alarm.
 
Additional Manufacturer Narrative
The biomedical engineer (bme) reported that the central nurse station (cns) did not alarm that a patient's spo2 readings were low and the patient coded, but they did not expire.The end users stated they did not silence or turn off the spo2 alarm.\ \ nihon kohden continues to investigate the reported event.Nihon kohden will submit a supplemental report in accordance with 21 cfr section 803.56 when additional information becomes available.\ \ \ additional device information: \ d10 concomitant medical device: the following device was used in conjunction with the cns:\ g7 monitor.\ model #: cu-172r.\ serial #:(b)(6).\ device manufacturer data: 5/25/2023.\ unique identifier (udi) #: (b)(4).\ returned to nihon kohden: not returned.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CNS-2101 Central Monitor
Common Device Name
Central Monitoring System
Manufacturer (Section D)
Nihon Kohden Corporation
attn: shama mooman
1-31-4 nishiochia
shinjuku-ku, tokyo 161-8 560
JA  161-8560
Manufacturer (Section G)
Nihon Kohden Tomioka Corporation
attn: shama mooman
1-1 tajino
tomioka city, gunma 370-2 314
JA   370-2314
Manufacturer Contact
shama mooman
seibu bldg 2, 4th floor 1-11-2
kusunokidai tokorozawa, saitama 359-8-580
JA   359-8580
MDR Report Key50159578
Report Number8030229-2026-06890
Device Sequence Number16096273
Product Code MHX
UDI-Device Identifier04931921017159
UDI-Public(01)04931921017159
Combination Product (Y/N)N
Initial Reporter StateAZ
Initial Reporter CountryUS
PMA/510(K) Number
K223567
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source User Facility
Initial Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date (Section B) 08/06/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberCNS-2101
Device Catalogue NumberCNS-2101
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/13/2026
Initial Report FDA Received Date08/06/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured02/06/2024
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
G7 Monitor.
Outcome Attributed to Adverse Event Other; Life Threatening;
Patient SexUnknown
-
-