• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: Tandem Diabetes Care t:slim X2 Insulin Pump with Control-IQ+ Technology; Alternate Controller Enabled Insulin Infusion Pump

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Tandem Diabetes Care t:slim X2 Insulin Pump with Control-IQ+ Technology; Alternate Controller Enabled Insulin Infusion Pump Back to Search Results
Model Number 1002717
Medical Device Problem Code Use of Device Problem (1670)
Health Effect - Clinical Code Hyperglycemia (1905)
Date of Event 05/21/2026
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that pump battery depleted too quickly, causing pump shutdown and leading customer to stop using pump for about a month, and customer later confirmed she had been on pump when she went to emergency room due to this issue.Customer did not know exact date of er visit but indicated it had occurred withing last 2 months.Customer¿s blood glucose reached 406 mg/dl and received saline and insulin through iv, and treatment resolved issue without permanent damage.Tandem technical support educated the customer about turning pump back on when it shuts down and reviewed logs which indicated that pump battery depletion is within normal parameters.
 
Additional Manufacturer Narrative
In use at the time of the event: cgm model: unknown mobile os: android / android_galaxy s20 fe 5g / unknown / android 16.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
t:slim X2 Insulin Pump with Control-IQ+ Technology
Common Device Name
Alternate Controller Enabled Insulin Infusion Pump
Manufacturer (Section D)
Tandem Diabetes Care
12400 high bluff drive
san diego CA 92130
Manufacturer Contact
john connell
12400 high bluff drive
san diego, CA 92130
+1(858)224-6281
MDR Report Key50159851
Report Number3013756811-2026-141780
Device Sequence Number9278355
Product Code QFG
UDI-Device Identifier00850006613786
UDI-Public(01)00850006613786(11)220622
Combination Product (Y/N)N
Initial Reporter StateGA
Initial Reporter CountryUS
PMA/510(K) Number
K250792
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Consumer
Type of Report Initial
Report Date (Section B) 08/06/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Model Number1002717
Device Catalogue Number1006406
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/21/2026
Initial Report FDA Received Date08/06/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured06/21/2022
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age31 YR
Patient SexFemale
Patient Weight233 LB
-
-