It was reported that a catheter issue occurred.The patient presented for an st-elevation myocardial infarction (stemi) procedure involving two layers of existing stents.The 100% stenosed target lesion was located in the moderately tortuous and moderately calcified mid-right coronary artery.An opticross hd imaging catheter was advanced for intravascular ultrasound (ivus) examination of the target lesion.The catheter would not cross the lesion, and during the use of a second opticross catheter, the tip broke off.The detached tip remained within the patient and was secured in place by a stent.The procedure was completed using another of the same device.There were no patient complications, and the patient outcome was stable.However, within days, the patient returned with a stemi approximately 20 mm proximal to the location of the retained tip.The last information obtained was that the patient was stable.
|