| Model Number |
AFR-00001 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Thrombosis/Thrombus (4440)
|
| Date of Event |
07/10/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that during cardiac ablation procedure, the patient had a non-medtronic device that had been implanted for approximately four years.The physician attempted to deliver a pulsed field ablation lesion near or on top of the device, and the temperature spiked rapidly during ablation.Two additional lesions were delivered near the area without further drastic temperature changes.Intracardiac echocardiography imaging of the left atrium identified clot or endothelial tissue in the area of the device.A suction device was used to remove the tissue.The procedure was aborted and the patient was under general anesthesia.No further patient complications have been reported as a result of this event.
|
| |
|
Additional Manufacturer Narrative
|
|
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|